MR-based vaginal vault brachytherapy in endometrial cancer: a dosimetric study
- NM
Norbert Mészáros 1,2
- GF
Georgina Fröhlich 3,2
- JV
Júlia Vizkeleti 2
- ÁH
Ádám Heintz 4
- EL
Edina Lukács 5
- ZN
Zoltán Nagy Takácsi 6,2
1. Department of Clinical Oncology, Semmelweis University, Faculty of Medicine, Budapest, Hungary
2. Center of Radiotherapy, National Institute of Oncology, Budapest, Hungary
3. Department of Biophysics, Eötvös Loránd University, Budapest, Hungary
4. Center of Oncological Imaging and Invasive Diagnostics, National Institute of Oncology, Budapest, Hungary
5. Multidisciplinary Center of Gynecological Oncology, National Institute of Oncology, Budapest, Hungary
6. Department of Radiotherapy, Semmelweis University, Faculty of Medicine, Budapest, Hungary
Select one of your emails
You have multiple emails registered with Frontiers:
Notify me on publication
Please enter your email address:
If you already have an account, please login
You don't have a Frontiers account ? You can register here
Abstract
Purpose. As de-escalation advances in endometrial cancer, more patients receive vaginal vault brachytherapy (VBT) as their sole adjuvant treatment, making patient-specific planning important. We compared geometry-based, non-image-guided planning with MRI-based anatomy-guided planning on non-contrast 0.55-T magnetic resonance imaging (MRI). Methods. Twenty-seven consecutive patients were treated with high-dose-rate VBT (single-channel applicator; 5 cm active length; 0.5 cm prescription depth); 25 with a reliably delineable vaginal wall were evaluable. Five received 3×7 Gy alone and 22 received 3×5 Gy plus pelvic external beam radiotherapy (45 Gy/1.8 Gy with chemotherapy in 6, or 46 Gy/2 Gy in 16). Both the delivered geometry-based plan and a retrospectively re-optimized MRI-based plan were evaluated on the same MRI-defined, tissue-only vaginal-wall CTV (applicator lumen excluded). Paired differences are Hodges–Lehmann estimates with 95% CI (Wilcoxon signed-rank test). Results. Median tissue-only CTV volume was 17.9 cm³. The MRI-based plan substantially improved target coverage over the geometry-based plan: V100 90.1% vs. 83.8% (+7.8%, 95% CI 3.9 to 12.9; p < 0.001) and D90 100.1% vs. 87.1% (+15.5%; p < 0.001), with higher EQD2-D90 (+3.0 Gy; p = 0.016). Geometry-based V100 fell below 90% in 21 of 25 patients. The MRI-based plan was more conformal (COIN +0.09; p < 0.001) and reduced rectal D2cc (−4.8 Gy; p < 0.001); bladder and sigmoid D2cc did not differ. Dog-ear grade correlated with reduced target dose (ρ = −0.44; p = 0.028). Conclusion. On the tissue-only vaginal-wall CTV, MRI-based anatomy-guided planning improved target coverage and conformality and reduced rectal dose, whereas the geometry-based protocol frequently undercovered the contoured vaginal-wall target, particularly where apical redundancy (dog ears) was present. These dosimetric findings, from a small single-center cohort, suggest a potential advantage of low-field MRI-based planning that requires confirmation in larger studies; the use of cervical cancer-derived OAR benchmarks and the absence of clinical endpoints preclude conclusions regarding clinical benefit.
Summary
Keywords
adjuvant radiotherapy, Endometrial cancer, image-based brachytherapy, low-field MRI, vaginal vault brachytherapy
Received
15 September 2026
Accepted
12 November 2026
Copyright
© 2026 Mészáros, Fröhlich, Vizkeleti, Heintz, Lukács and Takácsi. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Norbert Mészáros
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.